STUDY ON DETERMINE THE AMOUNT OF TRACE METALS Se, Cu, Zn, Mg, Mn IN THE BLOOD OF SOME RESPIRATORY PATIENTS BY METHOD OF INDUCTIVELY COUPLED PLASMA – MASS SPECTROMETRY (ICP – MS)

  • Trương Thị Bình Giang
  • Đinh Thị Trường Giang
Keywords: ICP-MS, blood, respiratory

Abstract

Concentration of selenium (Se), copper (Cu), zinc (Zn), magnesium (Mg), manganese (Mn) in blood of
patients with respiratory disease were determined by method of inductively Coupled Plasma - Mass
Spectrometry (ICP-MS) after sample digestionin a close container in a microwareoven. The result of Se, Cu,
Zn, Mg and Mn (ppm) are in the range of 0.0122 – 2.0865, 0.2329 – 1.5815, 1.9094 – 13.0964, 4.3917 –
19.1262 and 0.0430 – 0.5703. The detection limits are 0.3870 (ppt) for Se and 0.003; 0.045; 0.0042; 0.0045
(ppt) for Cu, Zn, Mg, Mn respectively. The relative standard deviation (%RSD) of 10 replicated
measurement is less than 6%. The recovery range of element is 50,83% to 108,35%. The accuracy of the
result shown in the repeatability and recovery performance is ensured according to AOAC.