DETERMINATION OF ZINC ISOTOPES IN BIOCHEMICAL SAMPLES BY ICP-MS PLASMA-INDUCED HIGH-FREQUENCY MASS SPECTROMETRY

  • Dung Nguyễn
  • Hà Nguyễn
  • Thảo Tạ
Keywords: ICP-MS, Zinc isotopes, isotopes analysis.

Abstract

In this study, a method of determining zinc isotopes (64Zn, 66Zn, 67Zn, 68Zn, 70Zn) in biochemical samples
by ICPMS inductive plasma mass spectrometry method is mentioned in this study. Analytical samples
were treated with a microwave wet inorganic technique for serum samples, urine samples were
centrifuged and then using ion-exchange chromatography were cleaned and removed influences. before
analysis. Factors affecting zinc isotope analysis were investigated, Plasma high-frequency power 1400W,
ion lens potential 9V for 5 individual isotopes, carrier gas flow for 2 isotopes 66Zn and 68Zn is 0, 0. 6L/min
and 64Zn, 67Zn and 70Zn are 0.9L/min. Analytical procedures were applied to identify 5 zinc isotopes 64Zn,
66Zn, 67Zn, 68Zn, 70Zn in the serum and urine samples. The results showed that the 70Zn isotope has a low
percentage in nature. Relative error meets the requirement of trace analysis.

điểm /   đánh giá
Published
2023-02-20